SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z540-1-1994

Some items we calibrate include temperature controllers, ovens, chart recorders, infrared guns, freezers, temperature and humidity devices, and handheld thermometers.

In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to our laboratory to perform the following Thermodynamic calibrations.

Please see full details below or on our full Scope of Accreditation (PDF) or learn more about our in-lab calibration service.

View our Electronic Test Stand

 

Thermodynamic Calibration Scope Includes:

Parameter/Equipment

Range CMC2,6 (±) Comments

Relative Humidity – Measure3

Hygrometers, Chart Recorders, Thermo- Hygrometers, Data Loggers 3

 

(10 to 95) % RH

 

3.2 % RH

 

Method: ACP-037; Veriteq SP2000-20R, Red Lion PAX panel meter

Temperature Measure – Chart Recorders, Thermo- Hygrometers, Data Loggers, Controllers, Thermometers3

(-40 to 100) °C
(> 100 to 300) °C
(> 300 to 1205) °C
0.08 °C
0.13 °C
0.9 °C

Method: ACP-037; Veriteq SP2000-20R, 7510DMM with RTD probe

Infrared Temperature – Measuring Equipment3

λ = (8 to 14) μm

(35 to 100) °C
(100 to 200) °C
(200 to 350) °C
(350 to 500) °C

(500 to 1200) °C

0.6 °C
0.66 °C
0.86 °C
1.2 °C

2.0°C

Method: ACP-037; Fluke 4181 precision infrared calibrator
ε = 0.95

Pegasus R ε = 0.995

Ovens, Furnaces, Chambers & Freezers3

 

(-30 to 0) °C

(0 to 1200) °C

 

1.4 °C

1.7 °C

Method: ACP-037 ASTM E145; uniformity

Keithley DMM7510 with RTD probeOmega DAQ56 data acquisition system & thermocouples

Have a need for Thermodynamic calibration products or services in Canton, Akron, Cleveland, Columbus or throughout Ohio?

CONTACT US

 

Already using our service?

LOGIN TO RETREIVE CALIBRATION CERTIFICATES

 

See our Entire Scope of Accreditation: 

A2LA Accredited Lab 

Certificate Number: 1848.01


1 This laboratory offers commercial calibration & field calibration services.

2 Calibration & Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device & to influences from the circumstances of the specific calibration.

3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibration & for other possible adverse effects such as those caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must also be considered & this, on its own, could result in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC.

4 In the statement of CMC, percentages are percentage of reading, unless otherwise indicated, L is the numerical value of the nominal length of the device measured in inches, R is the numerical value of the resolution of the device in microinches, D is the numerical value of the nominal diameter of the device measured in inches, I.V. represents “Indicated Value”, & Ra is the numerical value of the nominal roughness of the surface measured in micrometers roughness.

5 The stated measured values are determined using the indicated instrument (see Comments). This capability is suitable for the calibration of the devices intended to measure or generate the measured value in the ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a fraction/percentage of the reading plus a fixed floor specification.

6 The type of instrument or material being calibrated is defined by the parameter. This indicates the laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated for the listed measurement parameter.

7 This scope meets A2LA’s P112 Flexible Scope Policy.