SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z540-1-1994
In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to our laboratory to perform the following Chemical calibrations: pH meters and conductivity (fixed points).
Some companies we do work for include Ak Steel, Ashland Inc, JLG, Sherwin Williams, Federal Mogul, Park Ohio, Heinz, Goodyear, TRW.
Full details are listed below. You can also search all of our accredited calibrations or learn more about our in-lab calibration service.
Chemical Calibration Scope Includes:
| Parameter/Equipment | Range | CMC2, 4 (±) | Comments |
|---|---|---|---|
| pH Meter – Fixed Points3 | 4 pH | 0.12 pH | Method: ACP-071; pH buffer solutions |
| 7 pH | 0.12 pH | ||
| 10 pH | 0.12 pH | ||
| Conductivity – Fixed Points | (0 to 1) μS | 4.8 % | Method: ACP-071; conductivity standards |
| (1 to 10) μS | 1.2 % | ||
| (10 to 100) μS | 1.0 % | ||
| (100 to 1000) μS | 0.6 % | ||
| (1000 to 10 000) μS | 0.6 % | ||
| (10 000 to 100 000) μS | 0.5 % |
Notes from the A2LA scope:
- 2 CMC (Calibration & Measurement Capability) is the smallest uncertainty the lab can achieve within its scope for routine calibrations of nearly ideal equipment, expressed at approximately 95 % confidence (k = 2). The actual uncertainty of a specific calibration may be larger depending on the customer’s device and conditions.
- 3 Field (on-site) calibration service is available for this calibration. Uncertainties achievable at a customer’s site are normally expected to be larger than the CMC listed.
- 4 Percentages are percentage of reading unless otherwise indicated. L = nominal length in inches, D = nominal diameter in inches, I.V. = Indicated Value.
Source: A2LA Scope of Accreditation, Certificate No. 1848.01 (Revised 09/11/2025). The official PDF is the controlling document.